GOSSEN Foto- und Lichtmesstechnik GmbH 4
In order to enable sphere controller port click the box "Sphere controller settings" and
then type in the proper COM port number.
NOTE: If you do not know the port number of COM port, run Windows Control Panel/
Devices and Printers and check the USB <-> Serial COM port No.
Make sure that the dark current calibration has been done.
NOTE: When making dark current calibration, cover the entrance aperture in the sphere
to cut off the ambient light.